Columns

Bioinert columns
Accura Triart

Accura Triart is characterised by a bioinert surface coating on the column body and frit. It shows excellent peak shapes for highly adsorptive substances. Additionally, it prevents carry over and ensures complete recovery without column preconditioning. Therefore, it is especially ideal for the use in highly sensitive LC/MS analyses.
Accura Triart is suitable for analysis of nucleotides, proteins/peptides and metal-coordinating compounds.

Features
  • YMC-Triart materials filled in a bioinert surface coated hardware
  • Exceptional peak shapes with high sensitivities
  • Excellent recoveries without column preconditioning
  • Superior reproducibility and no carry-over effects
  • Ideal for highly sensitive LC/MS analyses
  • Reduced adsorption to column hardware

    Accura hardware with its inert surface prevents adsorption of oligonucleotides using a range of different buffers. Accura hardware further provides significantly higher recovery and sensitivity that can not be achieved with regular stainless steel hardware even after 20 sample injections. Depending on eluent conditions, adsorption may occur on PEEK hardware. The ready-to-use Accura hardware ensures high recovery and reproducibility from the very first use.

    Eluent condition 1: 8 mM TEA*-200 mM HFIP**/methanol (82/18)

    *triethyl amine, **1,1,1,3,3,3 hexafluoro-2-propanol

    Eluent condition 2: 100 mM TEAA* (pH 8)/methanol (82/18)

    *triethylammonium acetate

    Column size 50 X 2.1 mmI.D.
    Flow rate 0.42 mL/min
    Detection UV at 260 nm
    Sample All PS RNA 20mer

    No preconditioning for reliable results from the 1st injection

    Accura offers high recovery and reproducible peak areas from the first injection, unlike stainless steel column, which shows significant analyte adsorption even after repeated injections. By minimizing adsorption, Accura Triart eliminates preconditioning and enables more efficient RNA analysis.

    Column YMC-Triart Bio C4 (3 μm, 30 nm)
    100 X 2.1 mmI.D.
    Eluent A) 50 mM TEAA (pH 7.0)/acetonitrile (95/5)
    B) 50 mM TEAA (pH 7.0)/acetonitrile (50/50)
    9–14%B (0–10 min)
    Flow rate 0.2 mL/min
    Temperature 80℃
    Detection UV at 254 nm
    Sample Century™-Plus RNA Markers

    Enable low-adsorption and high-sensitivity analysis for a wide range of analytes

    Accura Triart provides high peak heights for oligonucleotides, peptides, and tryptophan metabolites compared to the regular stainless steel column. Accura Triart enhances the sensitivity significantly and helps to save precious samples without any loss.

    Oligonucleotides

    1. All PS RNA 20mer
    2. All PS RNA 21mer

    Column YMC-Triart Bio C18 (1.9 μm, 30 nm)
    50 X 2.1 mmI.D.
    Eluent A) 15 mM TEA-400 mM HFIP
    B) methanol
    8–18%B (0–10 min)
    Flow rate 0.42 mL/min
    Temperature 65°C
    Detection UV at 260 nm

    GLP‑1 receptor agonists

    1. Exenatide
    2. Semaglutide
    3. Retatrutide
    4. Tirzepatide
    5. Liraglutide

    Column YMC-Triart Bio C18 (1.9 μm, 30 nm)
    100 X 2.1 mmI.D.
    Eluent A) water/formic acid (100/0.1)
    B) acetonitrile/formic acid (100/0.1)
    30–60%B (0–6 min), 90%B (6–12 min)
    Flow rate 0.4 mL/min
    Temperature 60°C
    Detection MS (ESI+)

    Tryptophan metabolites

    1. Quinolinic acid
    2. Picolinic acid
    3. Kynurenine
    4. Xanthurenic acid
    5. Kynurenic acid
    6. Tryptophan

    Column YMC-Triart C18 (1.9 μm, 12 nm)
    50 X 2.1 mmI.D.
    Eluent A) 10 mM ammonium formate–formic acid (pH 3.7)
    B) acetonitrile
    5–50%B (0–3 min)
    Flow rate 0.4 mL/min
    Temperature 40°C
    Detection UV at 254 nm