Reversed Phase(ODS)

YMC-UltraHT Series

Columns


Characteristics of 2µm packing material

YMC-UltraHT series shows the same efficiency as sub-2µm columns for ultra-fast LC system and reduces the system pressure significantly.

Van Deemter curves/Column pressure

Column : 50 X 2.0 or 2.1 mmI.D.
Eluent : acetonitrile/water (60/40)
Temperature : 25ºC
Sample : Butyl benzoate

page top

Reduced analysis time and cost

This example shows the method transfer from conventional HPLC method to Ultra-fast LC method using a YMC-UltraHT column. Since the column efficiency is not decreased even when the linear velocity is 3 times as fast as the conventional condition, it can shorten analysis time by 90% without losing the separation characteristics.

Reduced analysis time and cost

Flavonoids

  1. Myricetin
  2. Quercetin
  3. Apigenin
  4. Kaempferol
  5. Baicalein
  6. Chrysin
  7. Acacetin
Column : acetonitrile/water/formic acid (35/65/0.1)
Temperature : 40ºC
Detection : UV at 260 nm

page top

Selecting the desired column dimensions

Feature and Use For LC/MS or reduction of solvent usage
(2.0 mmI.D.)
For conventional LC system
(3.0 mmI.D.)
Emphasize reduction of analysis time
(length: 30 mm)
30 X 2.0 -
Ultra-fast separation with the similar property as traditional column.
(length: 50 mm, 75 mm)
Standard for Ultra-fast LC
50 X 2.0
50 X 3.0
75 X 2.0 75 X 3.0
High resolution column for closely eluting peaks
(length: 100 mm)
100 X 2.0 100 X 3.0

page top

Ideal for High-resolution analysis: 100mm column

YMC-UltraHT column can resolve 38% more peaks than a conventional 5µm column with increased sensitivity.

Peptide mapping

Peptide mapping

Eluent : A) water/TFA(100/0.1), B) acetonitrile/TFA(100/0.1)
  5-50%B(0-15 min)
Flow rate : 0.4 mL/min
Temperature : 37ºC
Detection : UV at 220 nm
Injection : 1µL
Sample : Tryptic digest of β -Lactogloblin B (5 mg/mL; 37 ºC, 24 hr)

page top

Easily applicable for conventional LC system: 3.0mmI.D.

YMC-UltraHT series can be used for fast analysis with conventional LC system, as its pressure is lower than commercial sub-2µm column.Considering the extra column band broadening, the column of 3.0 mmI.D.is more applicable for conventional LC system than the column of 2.0 mmI.D.

50 X 3.0 mmI.D. / 50 X 2.0 mmI.D.

Cephalosporin antibiotics

  1. Cephalexin
  2. Cefaclor
  3. Cephaloglycin
  4. Cephaloridine
Column : YMC-UltraHT Pro C18
Eluent : acetonitrile/20mM KH2PO4 (10/90)
Temperature : 37ºC
Detection : UV at 260 nm

page top